Nitrous Oxide Sedation System S8800A
Application
Three level alarming system, visual and sound alarm information.
With lots of alarming, reminding and protection functions.
Advanced power management control technology.
With built-in backup power source, when outside power source goes off, back-up power source starts to work.
Ybo N2O Sedation Systemmakes a good performance in gastrointestinal endoscope department, obstetrics department, dentistry department and other departments. From high to low acuity, simple to complex cases, pediatric to geriatric patients, Ybo N2O Sedation Systems offer you the choices you need in ventilation, monitoring and technique. what’s more, our expertise in anesthesia delivery and ventilation is based on rich 23-year history of developing anesthesia solutions that meet your needs. Good performance based on High Security, High Accuracy, High Stability and accurate monitoring.
Safety
SPECIFICATIONS FOR S8800A NITROUS OXIDE SEDATION SYSTEM
|
|||
Ventilation modes
|
Continuous gas supply, On-demand gas supply, Rapid oxygen supply, Manual
|
||
Main parameters |
Oxygen concentration
|
30%-100%
|
|
N2O concentration
|
0%-70%
|
||
On-demand gas
|
0-30L/min
|
||
Pressure trigger(PTR)
|
-1cmH2O ~ -10cmH2O
|
||
Gas supply time
|
1s-6s
|
||
Peak flow
|
≥60L/min
|
||
Manual ventilation flow
|
0-30L/min
|
||
Rapid oxygen flow
|
25L/min-75L/min
|
||
Pressure limit
|
5cmH2O-60cmH2O
|
||
Monitoring parameters |
Oxygen concentration
|
30%-100%
|
|
Airway pressure
|
0-60cmH2O
|
||
Flow
|
0-30L/min
|
||
Graphical display
|
P-T(pressure – time)
|
||
Alarm and protection |
The AC power failure alarm
|
Power failure or no connection
|
|
Internal battery backup low voltage alarm
|
<11.3±0.3V
|
||
High oxygen concentration alarm
|
31%-100%
|
||
Low oxygen concentration alarm
|
30%-99%
|
||
High Airway pressure alarm
|
5cmH2O-60cmH2O
|
||
Pressure limit
|
60cmH2O
|
||
Emergency air input
|
When machine doesn’t work
|
||
N2O Closure function
|
When oxygen pressure ≤200kPa
|
||
Low gas source input pressure
|
Pressure ≤200kPa
|
||
Working conditions |
Gas source
|
O2,N2O
|
|
Pressure
|
280kPa-600kPa
|
||
Voltage
|
-220V±22V
|
||
Power frequency
|
50Hz±1Hz
|
||
Input power
|
60VA
|
||
Parking size
|