Veterinary Anesthesia Machine DM6B
Application
Three level alarming system, visual and sound alarm information.
With lots of alarming, reminding and protection functions.
Advanced power management control technology.
With built-in backup power source, when outside power source goes off, back-up power source starts to work.
The Veterinary Anesthesia machine makes a good performance in animal clinics. From high to low acuity, simple to complex cases, small to big animals, YBO Medical Anesthesia Systems offer you the choices you need in ventilation, monitoring and technique. what’s more, our expertise in anesthesia delivery and ventilation is based on rich 23-year history of developing anesthesia solutions that meet your needs. Good performance based on High Security, High Accuracy, High Stability and accurate monitoring.
SafetySpecification
Ventilation modes | DM6B |
Ventilator parameter ranges | |
Tidal volume(Vt) | 0, 20 mL ~ 1500 mL |
Frequence(Freq) | 1bpm ~ 100 bpm |
I/E | 2:1~1:6 |
Pressure Limit | 20 cmH2O ~ 100 cmH2O |
Monitored Parameters | |
Frequence(Freq) | 0 /min ~ 100 /min |
Tidal volume(Vt) | 0 mL ~ 2000 mL |
MV | 0 L/min ~ 100 L/min |
Graphical display: | |
P-T(pressure – time) | |
F-T(flow – time) | |
Size | |
1. Wooden case packing size:L 810*W 760*H 520mm , GW:53KG ; NW:30KG 2. Wooden case packing size:L 650*W 690*H 520mm , GW:40KG ; NW:20KG | |
Alarm and protection | |
The AC power failure alarm | Power failure or no connection |
Internal battery backup low voltage alarm | <11.3±0.3V |
No tidal volume | ≤5Ml within 6s |
High oxygen concentration alarm Low oxygen concentration alarm | 19%-100% 18%-99% |
High Airway pressure alarm Low Airway pressure alarm High Minute Volume alarm Low Minute Volume alarm Continuous Pressure alarm | 20cmH2O-100cmH2O 0cmH2O-20cmH2O Adult(5L/min-20L/min) Paed(1L/min-15L/min 0-10L/min) (PEEP+1.5kPa) over 16s |
Suffocation warning | 5s-60s no spontaneous ventilation |
The maximum limited pressure | <12.5 kPa |
Fan error Oxygen deficit | Show on screen Show on screen |
Working conditions | |
Gas source | O2 |
Pressure | 280kPa-600kPa |
Voltage | 100-240V |
Power frequency | 50/60Hz |
Input power | 60VA |